Bd Phoenix Automated Microbiology System-Minocycline (Gp) 1-32 Ug/Ml

by Becton, Dickinson & CO

510(k) K070809

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 9 Oct 2026. What this means

Regulatory status

as of registry sync 9 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK070809 Date24 Sep 2007ClearedStatusActive

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-0232-2023 FDA Class II initiated 13 Oct 2022

    Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic

    • Terminated Z-0570-2013 FDA Class II initiated 6 Nov 2012

    An antibiotic susceptibility test kit may produce false intermediate or false resistant results with clinically significant bacteria.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LON
Device Class
Class II
Regulation
21 CFR Part 866.1645
Advisory Committee
Microbiology
FDA Decision
SESE Substantially Equivalent