Behring Coagulation System
by Behring Diagnostics, Inc. 510(k)
Active
510(k) K970431
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK970431 | Date5 May 1997 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1435-2021
Siemens Healthcare Diagnostics Products GmbH has confirmed that the BCS XP System may be affected by a potential Emicizumab carryover on patient samples.
-
- Terminated Z-0825-2016
False short clotting times for PT on BCS and BCS XP with Dade Innovin can occur. This has also an effect for false low PT INR values. These false short clotting times were observed only with turbid, hemolytic and/or icteric samples from intensive care patients. The corresponding reaction curves of these samples showed a pre-peak, which can lead to falsely short clotting time. The result was not flagged by the existing check algorithms.
Specification
- 510(k) Number
- FDA Product Code
- JPA
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.5425
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent