Recalls

2 recalls · FDA

    • Terminated Z-1435-2021 FDA Class II initiated 2 Mar 2021

    Siemens Healthcare Diagnostics Products GmbH has confirmed that the BCS XP System may be affected by a potential Emicizumab carryover on patient samples.

    • Terminated Z-0825-2016 FDA Class II initiated 14 Jan 2016

    False short clotting times for PT on BCS and BCS XP with Dade Innovin can occur. This has also an effect for false low PT INR values. These false short clotting times were observed only with turbid, hemolytic and/or icteric samples from intensive care patients. The corresponding reaction curves of these samples showed a pre-peak, which can lead to falsely short clotting time. The result was not flagged by the existing check algorithms.