bellavista 302.323.000-40
Listed
Valid until
Not applicable
FDA recalls
None ongoingDocuments
None
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00198843002920 | Date22 Jul 2026 | StatusListed |
Recalls
-
- Completed Z-1967-2023
- Completed Z-1966-2023
Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.
-
- Terminated Z-0622-2022
Potential cessation in ventilation can occur under specific conditions.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade Name
- bellavista
- Generic Name
- Respiration monitor
- Model / Reference
- 302.323.000-40
- Primary DI / UDI-DI
- 00198843002920
- 510(k) Number
- FDA Product Code
- CBK
- GMDN Term
- Respiration monitor
- Device Class
- Class II
- Regulation Number
- 868.5895
- GUDID distribution status
- In Commercial Distribution
- MRI Safety
- MR Unsafe