Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 1000 ml/min); Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 750 ml/min)
by Belmont Medical Technologies
510(k) K242735
Active
510(k)
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK242735 | Date23 Dec 2024 | StatusActive |
Recalls
-
- Ongoing Z-1678-2025
Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.
-
- Ongoing Z-1099-2025
Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will need to be replaced by another set
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FRN
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5725
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent