Better-Bladder BB14
by Circulatory Technology, Inc.
Better-Bladder BB14 is a cardiotomy reservoir (FDA Class II; FDA product code DTN, Reservoir, Blood, Cardiopulmonary Bypass) made by Circulatory Technology, Inc., cleared under FDA 510(k) K981284 on 12 Aug 1998 and listed in FDA GUDID since 3 Oct 2016 (UDI-DI 00851997007004), active as of 22 Oct 2026. 7 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00851997007004 | Date3 Oct 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK981284 | Date12 Aug 1998 | StatusActive |
Official records
- AccessGUDID record 00851997007004
- FDA 510(k) database K981284
Across registries
- FDA 510(k) K981284 decided 12 Aug 1998 Active
- FDA GUDID UDI-DI 00851997007004 listed 3 Oct 2016 Listed
Recalls
-
- Ongoing Z-1862-2021
- Terminated Z-1171-2018
- Terminated Z-0817-2018
The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.
-
- Terminated Z-2832-2016
- Terminated Z-2050-2016
The seal between the balloon and the housing may leak.
-
- Terminated Z-0057-2017
Twisted or wrinkled balloon components which are considered a cosmetic defect by the firm.
-
- Terminated Z-1323-2010
During an FDA inspection, it was discovered that the firm had conducted a silent recall of its Bigger Better Bladder device after receiving a complaint from a hospital that the device collapsed during a procedure thereby obstructing flow.
Timeline
- Cleared FDA 510(k) K981284
- Recall initiated FDA recalls Z-1323-2010
- Recall initiated FDA recalls Z-2050-2016
- Recall initiated FDA recalls Z-2832-2016
- Recall initiated FDA recalls Z-0057-2017
- Listed FDA GUDID 00851997007004
- Recall initiated FDA recalls Z-1171-2018
- Recall initiated FDA recalls Z-0817-2018
- Recall initiated FDA recalls Z-1862-2021
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Better-Bladder
- Generic name
- Cardiotomy reservoir
- Model / Ref / Catalog
- BB14
- Primary DI / UDI-DI
- 00851997007004
- 510(k) Number
- FDA product code
- DTN
- GMDN term
- Cardiotomy reservoir
- Device class
- Class II
- Regulation number
- 870.4400
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Ethylene Oxide
- Size
- Lumen/Inner Diameter — 0.25 Inch
Similar devices
Other devices filed under FDA product code DTN (RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS).
- CARDIOTOMY RESERVOIR by Shiley, Inc.
- VENOMIDICARD PEDIATRIC VENOUS RESERVOIR WITH INTEGRALCARDIOTOMY FILTER/MIDICARD PEDIATRIC CARDIOTOMY RESERVOIR WITH FILT by Sorin Biomedical, Inc.
- CARDIOTOMY RESERVOIR W/FILTER BCR-3500 by American Bentley
- SCIMED R30 SILICONE ASSIST RESERVOIR by Scimed Life Systems, Inc.
- XTRA SEQUESTRATION SET X by SORIN GROUP ITALIA SRL
- HEMOBAG, MODEL HBBSD2000 by GLOBAL BLOOD RESOURCES LLC
- MODEL 4815 ADAPTER by Pace Medical
- MAXIMA* COLLAPSIBLE VENOUS RESERVOIR BAG & HOLDER by Johnson & Johnson Professionals, Inc.
- CAPIOX SX HARDSHELL RESERVOIR by TERUMO MEDICAL CORPORATION
- PLEUR-EVAC CARDIOTOMY RESERVOIR by Deknatel, Inc.