Bio-Modular® Shoulder System 113760
Bio-Modular Shoulder System 113760 is a device of GMDN type stem-fixed metallic humeral head prosthesis (FDA Class II) made by Biomet Orthopedics, Inc., cleared under FDA 510(k) K030710 on 4 Jun 2003 (2 earlier submissions) and listed in FDA GUDID since 23 Jul 2018 (UDI-DI 00887868253614), active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00887868253614 | Date23 Jul 2018 | StatusListed |
| United States | FDA 510(k)SESE | NumberK030710 | Date4 Jun 2003 | StatusActive |
Official records
- AccessGUDID record 00887868253614
- FDA 510(k) database K030710
- FDA 510(k) database K992119
- FDA 510(k) database K872454
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K030710 decided 4 Jun 2003 Active
- FDA GUDID UDI-DI 00887868253614 listed 23 Jul 2018 Listed
Recalls
-
- Terminated Z-1859-2017
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
-
- Terminated Z-0474-2013
The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear could potentially lead to a shorter implant life.
Timeline
- Cleared FDA 510(k) K872454
- Cleared FDA 510(k) K992119
- Cleared FDA 510(k) K030710
- Recall initiated FDA recalls Z-0474-2013
- Recall initiated FDA recalls Z-1859-2017
- Listed FDA GUDID 00887868253614
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Bio-Modular® Shoulder System
- Generic name
- Stem-fixed metallic humeral head prosthesis
- Model / Ref / Catalog
- 113760
- Primary DI / UDI-DI
- 00887868253614
- FDA product code
- HSD, KWT, KWS, MBF
- GMDN term
- Stem-fixed metallic humeral head prosthesis
- Device class
- Class II
- Regulation number
- 888.3690, 888.3650, 888.3660, 888.3670
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Stem-fixed metallic humeral head prosthesis.
- Anatomical Shoulder™ Domelock® 01.04213.425 by Zimmer, Inc.
- Anatomical Shoulder™ Domelock® 01.04213.465 by Zimmer, Inc.
- DELTA XTEND 130748021 by DEPUY (IRELAND)
- Humeris Shoulder 106-5220 by FX SOLUTIONS
- SMR Shoulder Resurfacing Head 1327.10.460 by Limacorporate
- Humelock II Cemented Shoulder System 106-5420 by FX SOLUTIONS
- SMR Shoulder Adaptor Taper 1330.15.276 by Limacorporate
- DELTA XTEND 130752026 by DEPUY (IRELAND)