Biolox Delta Ceramic Femoral Head
by Zimmer, Inc.
510(k) K071535
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK071535 | Date19 Nov 2007 | StatusActive |
Recalls
-
- Terminated Z-0336-2020
Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
-
- Terminated Z-1134-2015
Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.
-
- Terminated Z-0211-2013
Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy between product patient labels and product labels with respect to the reference and lot numbers of the product. While the information printed on the product label is correct, the info
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LZO
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3353
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent