Skip to content

Biomet Knee System 141336

by Biomet Orthopedics, Inc.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

7 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source GUDID, synced 23 Oct 2026. What this means

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00880304554375 UDI-DI Date24 Oct 2015ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK142933 Date23 Dec 2014DecidedStatusActive

Recalls

7 recalls · FDA

    • Terminated Z-2177-2018 FDA Class II initiated 18 May 2018

    One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented.

    • Terminated Z-1219-2017 FDA Class II initiated 15 Dec 2016

    Zimmer Biomet is conducting a medical device recall for ARCOS & TPRLC broaches and rasps due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.

    • Terminated Z-1756-2015 FDA Class III initiated 16 Apr 2015

    Biomet hip joint, knee joint and shoulder joint replacement components device labeling contains incorrect GTIN (Global Trade Identification Number) codes.

    • Terminated Z-2104-2011 FDA Unclassified (pre-2012) initiated 3 Mar 2011

    The firm is recalling the product due to the product's packaging being labeled "I-Beam 75mm. Tibial Plate", but the package actually contains the "Interlok 67mm. CR Tibial Plate".

  • 2 recalls, same reason

    • Terminated Z-0706-2008 FDA Unclassified (pre-2012) initiated 12 Dec 2007
    • Terminated Z-0707-2008 FDA Unclassified (pre-2012) initiated 12 Dec 2007

    Mislabeled as to size: Package labeled as containing a 67 mm tray actually contains a 75 mm tray and vice versa.

    • Terminated Z-0255-2008 FDA Unclassified (pre-2012) initiated 18 Sep 2007

    Mislabeled as to size: the package actaully contained a 71 mm tray.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Biomet Knee System
Generic name
Uncoated knee tibia prosthesis, metallic
Model / Ref / Catalog
141336
Primary DI / UDI-DI
00880304554375
510(k) Number
FDA product code
MBH, JWH, MBV
GMDN term
Uncoated knee tibia prosthesis, metallic
Device class
Class II
Regulation number
888.3565, 888.3560
GUDID distribution status
In Commercial Distribution

Other devices from Biomet Orthopedics, Inc.

See all devices from Biomet Orthopedics, Inc. →

Properties

  • Supplied sterile
  • Single use
  • Implantable
  • Prescription use
  • Lot / batch number
  • Expiration date

Other properties (9)

  • Sterilise before use: No
  • Contains latex: No
  • Over-the-counter: No
  • Convenience kit: No
  • Combination product: No
  • HCT/P: No
  • Premarket exempt: No
  • Serial number: No
  • Manufacturing date: No