Bleas Esiriuc Anaesthsia System, Models 3000, 2000
by Blease USA, Inc. 510(k)
Active
510(k) K051629
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK051629 | Date23 Sep 2005 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2113-2010
Potential for the touch screen on the ventilator portion of the device to go blank when touched. Possibility that this failure may also cause mechanical ventilation to stop.
-
- Terminated Z-1408-2010
Potential for delivery of a lower than set concentration of anaesthetic to a patient due to a sticking valve. The lower shuttle may become jammed, not permitting flow of anaesthesia gas through the vapouriser. The vapouriser back bar mounting valves will require exchange only on the systems identified in this matter, no other systems are affected.
-
- Terminated Z-0813-2011
Users may experience difficulty turning the control knob to an active vapour delivery setting, or may not be able to turn on the Left Hand mounted vapouriser on the Blease Sirius Anaesthesia Systems.
Specification
- 510(k) Number
- FDA Product Code
- BSZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5160
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent