Skip to content

Blease Frontline Plus Range, Anesthesia Machines, Model 440, 560 and 690

by Blease Medical Equipment, Ltd. 510(k)

Active United States

510(k) K003251

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK003251 Date23 May 2001ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

3 recalls · FDA

    • Terminated Z-0352-2014 FDA Class I initiated 17 Oct 2013

    Spacelabs Healthcare Ltd. of United Kingdom recalls , and Service Kits, Part Number 050-0659-00 and 050-0901-00. The Bag to-Vent switch in CAS I/II Absorbers may fail due to loose fastening hardware in an absorber.

    • Terminated Z-0968-2013 FDA Class I initiated 15 Jan 2013 on shared 510(k) K003251

    A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.

    • Terminated Z-0565-2013 FDA Class II initiated 2 Nov 2012

    The BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators were shipped with manuals that describe a secondary O2 Supply Failure alarm which has not been incorporated into the system. The manuals were prematurely updated for a feature that is not yet available.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Specification

510(k)

510(k) Number
FDA Product Code
BSZ
Models / Variants
FDA applicant
Blease Medical Equipment , Ltd.
Device Class
Class II
Regulation
21 CFR Part 868.5160
Advisory Committee
Anesthesiology
FDA Decision
SESE Substantially Equivalent

Other devices from Blease Medical Equipment, Ltd.