Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its intended use. The most commonly reported complaints were related to impaired software functioning, impaired instrumentation functioning, and general difficulties using the product.
-
- Ongoing Z-0846-2025
The Pin Guide is being recalled because the metal tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset being unable to detect the Pin Guide and potentially prolonging surgery time or changing of surgical method from guided to conventional.