Bridge Occlusion Balloon
by Spectranetics Corporation 2 variants
510(k) K153530
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK153530 | Date5 Feb 2016 | StatusActive |
Recalls
-
- Terminated Z-2143-2020
There is a potential for thrombus formation on balloons prepped and staged in the vasculature for long durations. All users, such as those performing lead extraction cases may not be aware of this risk. It was determined that current labeling is insufficient to ensure all potential users minimize the time the balloon remains in vasculature. If left in too long it could lead to embolization.
-
- Terminated Z-3106-2017
Possible inability to pass the guidewire through the Bridge device, therefore, resulting in a potential delay of treatment.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MJN
- FDA applicant
- Spectranetics, Inc.
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4450
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent