# BrightMatter Guide with Surface Trace Registration

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/brightmatter-guide-with-surface-trace-registration-haw>
- Manufacturer: [Synaptive Medical Inc](https://devicerts.com/manufacturers/fda-fei-3012075008) (Canada)
- Device class: Class II (US FDA)
- FDA product code: HAW

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K153281 · decision 2016-03-30 · decision SESE · status: Active

## Recalls

- Z-0470-2019 · Class I · terminated · initiated 2018-10-09: This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.

## Identifiers

- 510(k) Number: K153281 [FDA 510(k)]
- FDA product code: HAW [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 882.4560 [FDA 510(k)]
- Advisory committee: Neurology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: BrightMatter Guide with Surface Trace Registration, Synaptive Medical Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/brightmatter-guide-with-surface-trace-registration-haw (data as of 2026-11-28).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-28
- JSON: <https://devicerts.com/devices/brightmatter-guide-with-surface-trace-registration-haw.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/brightmatter-guide-with-surface-trace-registration-haw
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.