  "format": "devicerts.device.v1",
  "cite_as": "Brightview Vct Imaging System, Philips Medical Systems (Cleveland), Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/brightview-vct-imaging-system-kps (data as of 2026-11-21).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/brightview-vct-imaging-system-kps",
    "json_url": "https://devicerts.com/devices/brightview-vct-imaging-system-kps.json",
    "markdown_url": "https://devicerts.com/devices/brightview-vct-imaging-system-kps.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-21",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/brightview-vct-imaging-system-kps",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Brightview Vct Imaging System",
    "registry_name": "BRIGHTVIEW VCT IMAGING SYSTEM",
    "url": "https://devicerts.com/devices/brightview-vct-imaging-system-kps",
    "manufacturer": {
      "name": "Philips Medical Systems (Cleveland), Inc.",
      "url": "https://devicerts.com/manufacturers/philips-medical-systems-cleveland-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K080927",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "KPS"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K080927",
      "number_label": null,
      "device_udi_di": null,
      "date": "2008-04-11",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0930-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-12-15",
      "reason": "Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-0931-2024",
      "classification": "Class I",
      "status": "ongoing",
      "initiated": "2023-12-15",
      "reason": "Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-0678-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-11-30",
      "reason": "While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or\ncontusion.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-0679-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-11-30",
      "reason": "While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or\ncontusion.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2009-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-06-10",
      "reason": "An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2011-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-06-10",
      "reason": "An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2010-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2019-06-10",
      "reason": "An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1753-2017",
      "classification": "Class III",
      "status": "terminated",
      "initiated": "2017-02-24",
      "reason": "The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1752-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-24",
      "reason": "The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1750-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-24",
      "reason": "The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1751-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-24",
      "reason": "The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1481-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-01-31",
      "reason": "Four issues: \n1. Motion controller problem stops scan and no data image produced. \n2. Door interlock switch problem disables CT scan. \n3. Detector contacts head holder when performing Patient Unload. \n4. JETStream freezes during gated planar scan",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1480-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-01-31",
      "reason": "Four issues: \n1. Motion controller problem stops scan and no data image produced. \n2. Door interlock switch problem disables CT scan. \n3. Detector contacts head holder when performing Patient Unload. \n4. JETStream freezes during gated planar scan",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1479-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-01-31",
      "reason": "Four issues: \n1. Motion controller problem stops scan and no data image produced. \n2. Door interlock switch problem disables CT scan. \n3. Detector contacts head holder when performing Patient Unload. \n4. JETStream freezes during gated planar scan",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1186-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-12-23",
      "reason": "The intended use listed in the English and localized language IFUs for BrightView XCT systems did not match the intended use documented in the 510K submission.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1187-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-12-23",
      "reason": "The intended use listed in the English and localized language IFUs for BrightView XCT systems did not match the intended use documented in the 510K submission.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2737-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-05-25",
      "reason": "The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient.  The issue was caused by a software defect in the workflow setup script.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2736-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-05-25",
      "reason": "The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient.  The issue was caused by a software defect in the workflow setup script.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1022-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-12-17",
      "reason": "The firm discovered four issues with the BrightView Family systems.\n\nIssue 1: \nDuring a cardiac non-Auto Body Contouring (ABC) scan, using Cardiac High Resolution (CHR) collimators, the collimator may come in contact with the bottom edge of the patient pallet.  If this occurs, collision sensors would pause the scan.\n\nIssue 2:\nDuring a scan, the gantry motion may halt prematurely.  This cause",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1792-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-03-06",
      "reason": "The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1011-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-11-12",
      "reason": "Unintended detector and gantry movement due to software issues.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-0450-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-11-04",
      "reason": "Software issues",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1982-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-06-23",
      "reason": "Flat Panel Display (FPD) failed to remain securely locked in the deployed position.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-1074-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-12-19",
      "reason": "Three problems were reported to the firm:\n\nWhen positioning for a relative 180 degree non circular acquisition the second (bottom) detector was not positioned correctly causing an inappropriate rotation orbit calculation.  When the acquisition was started and the detectors began to move into position, the incorrect calculation resulted in the detectors coming in contact with the patient.\n\t\nAs",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-0203-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-10-23",
      "reason": "It was discovered the CT portion of the scan was interrupted ( stopping the CT exposure at the time of interruption) and did not complete successfully.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-0089-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-09-03",
      "reason": "Philips received reports from the field that the Flat Panel Detector (FPD) does not always remain securely latched in the stowed position.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2948-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-06-15",
      "reason": "Non-conformances  with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2949-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-06-15",
      "reason": "Non-conformances  with BrightView X/XCT systems running version 2.0.1 including data transfer, XCT acquisition, Pixel Mask for image reconstruction and table moving incorrectly.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2155-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-03-28",
      "reason": "The current Headholder to support Bright View XCT brain imaging allowed for imaging the brain but doe allow for imaging of the associated shoulder/neck area of the body.",
      "applies": "this device",
      "via": "K080927"
    },
    {
      "number": "Z-2156-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-03-28",
      "reason": "The current Headholder to support Bright View XCT brain imaging allowed for imaging the brain but doe allow for imaging of the associated shoulder/neck area of the body.",
      "applies": "this device",
      "via": "K080927"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K080927",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "KPS",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1200",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}