  "format": "devicerts.device.v1",
  "cite_as": "Bubble Detector Sensor 1/4\", LivaNova Deutschland GmbH. deviCERTS medical device catalogue. https://devicerts.com/devices/bubble-detector-sensor-1-4 (data as of 2026-11-17).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/bubble-detector-sensor-1-4",
    "json_url": "https://devicerts.com/devices/bubble-detector-sensor-1-4.json",
    "markdown_url": "https://devicerts.com/devices/bubble-detector-sensor-1-4.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-23",
    "status_as_of": "2026-11-17",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/bubble-detector-sensor-1-4",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Bubble Detector Sensor 1/4\"",
    "registry_name": "Bubble Detector Sensor 1/4\"",
    "url": "https://devicerts.com/devices/bubble-detector-sensor-1-4",
    "manufacturer": {
      "name": "LivaNova Deutschland GmbH",
      "url": "https://devicerts.com/manufacturers/livanova-deutschland-gmbh",
      "country": "Germany",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "04033817900573",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K071318",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "DTQ"
    ],
    "emdn": [],
    "gmdn": [
      {
        "code": "35099",
        "term": "Cardiopulmonary bypass system, roller type"
      }
    ],
    "description": null,
    "noindex": true
  },
  "authorizations": [
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "04033817900573",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2015-10-20",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K071318",
      "number_label": null,
      "device_udi_di": null,
      "date": "2007-07-06",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1119-2020",
      "classification": "Class III",
      "status": "terminated",
      "initiated": "2019-01-28",
      "reason": "Incorrect labeling; it was identified that some Datapad devices contained incorrect labels. Specifically the manufacturing date in the production identifier was incorrect and the manufacturing date on the unique device identifier  label was labeled  in the format of yy-mm-dd instead of yyyy-mm-dd.",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-2866-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-08-25",
      "reason": "Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-0856-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-12-04",
      "reason": "Sorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob failures resulting in no response.",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-0142-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-09-24",
      "reason": "Sorin Group is recalling certain models of S5 Perfusion Systems due to malfunction of cardioplegia control.",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-0274-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-09-19",
      "reason": "Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a \"Fault in motor controller\" alarm message displayed on the system panel.  This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury.  An out of specification electrical component was determin",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-0275-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-09-19",
      "reason": "Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a \"Fault in motor controller\" alarm message displayed on the system panel.  This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury.  An out of specification electrical component was determin",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-0271-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-09-19",
      "reason": "Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the\nSorin Group S5 Perfusion System due to reports of a \"Fault in motor controller\" alarm message displayed on the system panel.  This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury.  An out of specification electrical component was determi",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-0272-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-09-19",
      "reason": "Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a \"Fault in motor controller\" alarm message displayed on the system panel.  This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury.  An out of specification electrical component was determin",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-0273-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-09-19",
      "reason": "Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a \"Fault in motor controller\" alarm message displayed on the system panel.  This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury.  An out of specification electrical component was determin",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-1325-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-02-23",
      "reason": "The pump speed control knob is attached to the shaft of an encoder that is used to adjust the speed. Insufficient lubrication within the encoder increases friction over time making it more difficult to turn the pump speed control knob. Excessive friction causes premature wear of the encoder shaft which can result in a stuck shaft.",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-1327-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-02-23",
      "reason": "The pump speed control knob is attached to the shaft of an encoder that is used to adjust the speed. Insufficient lubrication within the encoder increases friction over time making it more difficult to turn the pump speed control knob. Excessive friction causes premature wear of the encoder shaft which can result in a stuck shaft.",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-1328-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-02-23",
      "reason": "The pump speed control knob is attached to the shaft of an encoder that is used to adjust the speed. Insufficient lubrication within the encoder increases friction over time making it more difficult to turn the pump speed control knob. Excessive friction causes premature wear of the encoder shaft which can result in a stuck shaft.",
      "applies": "this device",
      "via": "K071318"
    },
    {
      "number": "Z-1326-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-02-23",
      "reason": "The pump speed control knob is attached to the shaft of an encoder that is used to adjust the speed. Insufficient lubrication within the encoder increases friction over time making it more difficult to turn the pump speed control knob. Excessive friction causes premature wear of the encoder shaft which can result in a stuck shaft.",
      "applies": "this device",
      "via": "K071318"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "Bubble Detector",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Cardiopulmonary bypass system, roller type",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Model / reference",
          "values": [
            {
              "value": "Sensor 1/4\"",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "04033817900573",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K071318",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "DTQ",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Cardiopulmonary bypass system, roller type",
              "code": "35099",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "870.4220",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}