# Buretrol Solution Sets

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/buretrol-solution-sets-fpa>
- Manufacturer: [Baxter Healthcare Corp](https://devicerts.com/manufacturers/baxter-healthcare-corp) (United States, FDA 510(k) applicant)
- Name in the registry: BURETROL SOLUTION SETS
- Device class: Class II (US FDA)
- FDA product code: FPA

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K984381 · decision 1999-02-19 · decision SESE · status: Active

## Recalls

- Z-0012-2013 · Class I · terminated · initiated 2012-09-07: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
- Z-0013-2013 · Class I · terminated · initiated 2012-09-07: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
- Z-0014-2013 · Class I · terminated · initiated 2012-09-07: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
- Z-0015-2013 · Class I · terminated · initiated 2012-09-07: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
- Z-0070-2011 ·  · terminated · initiated 2010-09-03: There are leaks in the Drip Chamber Check Valve Subassembly of the Solution Set. A leak in a sterile pathway could allow microbial entry, which could lead to a blood stream infection.
- Z-0487-2010 ·  · terminated · initiated 2009-11-03: The tubing may separate from the distal Clearlink Y-site of the set, which could result in patient injury if the separation occurs during patient use.
- Z-0654-06 ·  · terminated · initiated 2005-12-15 · matched through shared submission K984381, may concern other models: The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
- Z-0655-06 ·  · terminated · initiated 2005-12-15 · matched through shared submission K984381, may concern other models: The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
- Z-0656-06 ·  · terminated · initiated 2005-12-15 · matched through shared submission K984381, may concern other models: The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
- Z-0199-06 ·  · terminated · initiated 2005-03-17: The Buretrol solution sets are currently labeled as acceptable for use with the Colleague Volumertric Pump, but there is a potential for a non-detection of an upstream occlusion under certain conditions.
- Z-0845-05 ·  · terminated · initiated 2005-02-28: Possible occlusion of the Clearlink injection site valve due to inadequate silicone lubrication of the slit in the upper seal of the valve.
- Z-0193-05 ·  · terminated · initiated 2004-04-28: Inadequate directions for use may have contributed to an increase in infection rates at a small number of customers beginning use of the Clearlink needleless device.
- Z-0504-03 ·  · terminated · initiated 2002-10-17: The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.
- Z-0508-03 ·  · terminated · initiated 2002-10-17: The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.
- Z-0510-03 ·  · terminated · initiated 2002-10-17: The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.
- Z-0523-03 ·  · terminated · initiated 2002-10-17: The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.

## Identifiers

- 510(k) Number: K984381 [FDA 510(k)]
- FDA product code: FPA [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 880.5440 [FDA 510(k)]
- Advisory committee: General Hospital [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Buretrol Solution Sets, Baxter Healthcare Corp. deviCERTS medical device catalogue. https://devicerts.com/devices/buretrol-solution-sets-fpa (data as of 2026-11-24).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-24
- JSON: <https://devicerts.com/devices/buretrol-solution-sets-fpa.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/buretrol-solution-sets-fpa
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.