Skip to content

CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion Pump, Model 2110; PharmGuard® Administrator Medication Safety Software; CADD® Administration Sets; Wireless Communication Module (CM)

by Smiths Medical Ads, Inc.

510(k) K170982

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

6 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 11 Oct 2026. What this means

Regulatory status

as of registry sync 11 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK170982 Date24 Aug 2017ClearedStatusActive

Recalls

6 recalls · 6 ongoing · FDA

    • Ongoing Z-1667-2025 FDA Class I initiated 10 Apr 2025

    Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

    • Ongoing Z-1681-2025 FDA Class I initiated 10 Apr 2025

    There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

    • Ongoing Z-1662-2025 FDA Class I initiated 10 Apr 2025

    Pumps may experience Wireless Connection Modules intermittent connection alarms, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

    • Ongoing Z-0466-2025 FDA Class II initiated 26 Jun 2024

    The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.

    • Ongoing Z-2420-2024 FDA Class I initiated 27 Feb 2024

    The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.

    • Ongoing Z-0963-2023 FDA Class I initiated 9 Dec 2022 on shared 510(k) K170982

    There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
MEA
Device Class
Class II
Regulation
21 CFR Part 880.5725
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent