Carescape B850 E-musb
510(k) K213336
Active
Valid until
No date given
FDA recalls
1 ongoingTrust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE 510(k) | NumberK213336 | Date13 Apr 2022 | StatusActive |
| European Union | CE/MDR EUDAMED | Number— | Date— | StatusActive |
Official records
Across registries
- FDA 510(k) K213336 decided 13 Apr 2022 Active
- EUDAMED Active
Recalls
-
- Ongoing Z-2170-2024
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
Timeline
- Cleared FDA 510(k) K213336
- Recall initiated FDA recalls Z-2170-2024
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- 510(k) Number
- FDA product code
- MHX
- Device class
- Class II
- Regulation
- 21 CFR Part 870.1025
- Advisory committee
- Cardiovascular
- FDA decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 18 Sep 2026
The Carescape B850 E-musb is a physiological patient monitor designed for continuous cardiac and respiratory monitoring, featuring arrhythmia detection and alarm capabilities. It falls under the classification of a monitor for patients with arrhythmia detection or alarms, providing real-time vital sign assessment in clinical settings.
This device is intended for use in hospital, critical care, and emergency environments, supplied as a reusable medical instrument. It complies with FDA 510(k) clearance (K213336) and MDR authorisation, adhering to regulatory standards for Class II devices under product code MHX. No specific sterility or MRI safety information is provided in the data.