# Carescape R860

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/carescape-r860-cbk>
- Manufacturer: [Datex-Ohmeda Inc.](https://devicerts.com/manufacturers/datex-ohmeda-inc) (United States)
- Name in the registry: CARESCAPE R860
- Device class: Class II (US FDA)
- FDA product code: CBK

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K142679 · decision 2015-06-12 · decision SESE · status: Active

## Recalls

- Z-1227-2022 · Class I · ongoing · initiated 2022-04-18: Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.
- Z-1226-2022 · Class I · ongoing · initiated 2022-04-18: Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.
- Z-2883-2020 · Class II · terminated · initiated 2020-07-14: The oxygen sensor for certain CARESCAPE R860 devices has a potential issue that may result in an inaccurate display of FiO2 value from what is being delivered.
- Z-0463-2019 · Class I · terminated · initiated 2018-10-05: The ventilator Inspiratory Safety Guard (ISG) may disconnect from the breathing circuit pathway. As a result, this could create a loss of ventilation which may lead to inadequate oxygenation for patients, increasing the possibility of hypoxia.
- Z-2738-2018 · Class II · terminated · initiated 2018-06-28: The Neonatal Flow Sensor Cable may be missing the connector housing exposing the wires in the cable.
- Z-2266-2018 · Class II · terminated · initiated 2017-12-08: Potential for the display processor to experience an unexpected failure.

## Identifiers

- 510(k) Number: K142679 [FDA 510(k)]
- FDA product code: CBK [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 868.5895 [FDA 510(k)]
- Advisory committee: Anesthesiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Carescape R860, Datex-Ohmeda Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/carescape-r860-cbk (data as of 2026-11-19).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-19
- JSON: <https://devicerts.com/devices/carescape-r860-cbk.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.