Case
by GE Medical Systems Information Technologies, Inc.
Ref 2035123-001Model CASE V7.0
Active
CE MDR
Valid until
Not listed
FDA recalls
1 ongoingDocuments
None linked in EUDAMED
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing GUDID | Number00195278403643 | Date25 Mar 2025 | StatusListed |
| European Union | CE/MDR EUDAMED | Number8406821BUG00170H6 00195278403643 | Date2 Jun 2022 | StatusActive |
Certificates
EUDAMED lists no certificate for this device's Basic UDI-DI.
Recalls
-
- Ongoing Z-2279-2026
Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- Generic Name
- Electrocardiograph, professional, multichannel
- Model / Ref / Catalog
- CASE V7.0, 2035123-001, V7.0
- Primary DI / UDI-DI
- 00195278403643
- Basic UDI-DI
- 8406821BUG00170H6
- 510(k) Number
- FDA Product Code
- DQK
- EMDN Code
- Z12050301 General purpose electrocardiographs
- GMDN Term
- Electrocardiograph, professional, multichannel
- Device Class
- Class IIa EUDAMEDClass II GUDID
- Regulation Number
- 870.1425
- EU market of registration
- France
- GUDID distribution status
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | CASE V7.0 2035123-001 | V7.0 |
| Device Class | Class IIa | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000017529
- Authorised representative
- GE Medical Systems SCS FR-AR-000000344
Other devices from GE Medical Systems Information Technologies, Inc.
See all devices from GE Medical Systems Information Technologies, Inc. →
Properties
Other properties (20)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No