Cell-Dyn Sapphire System
by Abbott Laboratories 510(k)
Active
510(k) K051215
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK051215 | Date21 Jul 2005 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-0604-2024
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
-
- Terminated Z-2353-2010
Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks.
-
- Terminated Z-2421-2010
Under certain conditions in closed mode of operations, the Aspiration Probe Detector PCB assembly may fail to operate properly and provide inaccurate results.
-
- Terminated Z-1069-2009
Some syringes were improperly assembled and are therefore not usable, due to an incorrect nut on the syringe shell.
-
- Terminated Z-0306-2009
Exposed wire resulted in minor electrical shock and burn to the service technician
-
- Terminated Z-2139-2008
Table in LIS Specification depicting the association between Record ID and Numerical Result Label is incorrect. If a record ID is used to configure the system for mapping, results will come out nonsensical.
-
- Terminated Z-1165-2008
Count may be out of specification--Hemoglobin background count may be out of specification (high) after installation of new CELL-DYN Sapphire Hemoglobin Reagent Syringes with packaging dates between May 8, 2007 and November 29, 2007.
-
- Terminated Z-1103-2008
Microbial Contamination: The WBC (White Blood Cell) Reagent Part A was confirmed positive for Pseudomonas contamination.
-
- Terminated Z-0301-2008
Syringe Failure: Insufficient lubrication of the plunger tip may cause syringes to fail at installation or shortly thereafter
-
- Terminated Z-0565-2007
Results for tests of RBC, MCV, RDW, MPV and PLT may be falsely elevated or falsely decreased when the Cell Dyn Diluent/Sheath reagent is used with Cell-Dyn 4000, Ruby and Sapphire systems.
-
- Terminated Z-0708-2007
High platelet background count when using the CELL-DYN 4000 System and/or CELL-DYN Sapphire System; with patient results unacceptable -out-of-range.
Specification
- 510(k) Number
- FDA Product Code
- GKZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.5220
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent