Celsius ThermoCool DI7TCDLRT
Celsius ThermoCool DI7TCDLRT is a Class III device made by Biosense Webster Inc, approved under FDA PMA P030031 on 5 Nov 2004 and listed in FDA GUDID since 10 Sep 2014 (UDI-DI 10846835000443), active as of 23 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10846835000443 | Date10 Sep 2014 | StatusListed |
| United States | FDA PMAAPPR | NumberP030031 | Date5 Nov 2004 | StatusActive |
Official records
- AccessGUDID record 10846835000443
- FDA PMA database P030031
Across registries
- FDA PMA P030031 decided 5 Nov 2004 Active
- FDA GUDID UDI-DI 10846835000443 listed 10 Sep 2014 Listed
Recalls
-
- Ongoing Z-0169-2023
Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.
-
- Terminated Z-0139-2015
The recall was initiated because Biosense Webster is providing additional labeling for the safe and effective use of the ThermoCool SmartTouch Catheter.
-
- Terminated Z-0981-2012
- Terminated Z-0983-2012
- Terminated Z-0980-2012
- Terminated Z-0982-2012
- Terminated Z-0979-2012
Biosense Webster, Inc. has recently become aware of the omission of the following two Warning and Precaution statements on the Instructions For Use (IFU) for the above catheter: 1. If phrenic nerve location is a concern, precautionary measures are recommended to evaluate the proximity of the nerve to the ablation electrode, such as pacing maneuvers. 2. To minimize risk of atrio-esophageal fist
-
- Terminated Z-0057-06
Increased complaints regarding catheters with leaks and flow obstruction.
Timeline
- Approved FDA PMA P030031
- Recall initiated FDA recalls Z-0057-06
- Recall initiated FDA recalls Z-0981-2012
- Recall initiated FDA recalls Z-0983-2012
- Recall initiated FDA recalls Z-0980-2012
- Recall initiated FDA recalls Z-0982-2012
- Recall initiated FDA recalls Z-0979-2012
- Listed FDA GUDID 10846835000443
- Recall initiated FDA recalls Z-0139-2015
- Recall initiated FDA recalls Z-0169-2023
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade name
- Celsius ThermoCool
- Generic name
- Cardiac radio-frequency ablation system catheter
- Model / Ref / Catalog
- DI7TCDLRT
- Primary DI / UDI-DI
- 10846835000443
- PMA number
- FDA product code
- OAD
- GMDN term
- Cardiac radio-frequency ablation system catheter
- Device class
- Class III
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category