Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur Analyzer
510(k) K971418
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK971418 | Date17 Jul 1997 | StatusActive |
Recalls
-
- Terminated Z-1553-2014
Instrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling.
-
- Terminated Z-1119-2008
Incorrect results: False reactive specimens on the ADVIA Centaur HAV IgM assay.
-
- Terminated Z-0645-2007
Firm received complaints for signal 2 & 4 errors, and shifts in Relative Light Units (RLU). Investigations showed that the On-Board Stability (OBS) of the acid & base reagents may be compromised after 4 days after installation onto the Centaur CP system due to evaporation.
-
- Terminated Z-1516-06
Sample/Patient mis-identification (software defect)-the system can associate a test result and sample identification (SID) with an incorrect patient name when the Patient Demographics feature is used and the Patient Identification (PID) field is left blank
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MOI
- Device Class
- Class II
- Regulation
- 21 CFR Part 866.6010
- Advisory Committee
- Immunology
- FDA Decision
- SESE Substantially Equivalent