Cios Spin
by Siemens Medical Solutions USA, Inc. 2 variants
510(k) K181550
Cios Spin is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K181550 on 30 Oct 2018, active as of 22 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK181550 | Date30 Oct 2018 | StatusActive |
Official records
- FDA 510(k) database K181550
Recalls
-
- Terminated Z-2888-2020
Siemens has become aware of a hardware error which affects Cios Spin and Cios Alpha systems after the previous repair of the X-ray Generator with an integrated Energy Storage Unit (ESU).
-
- Terminated Z-2848-2020
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
-
- Terminated Z-1959-2020
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
-
- Terminated Z-2128-2020
The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.
Timeline
- Cleared FDA 510(k) K181550
- Recall initiated FDA recalls Z-2128-2020
- Recall initiated FDA recalls Z-1959-2020
- Recall initiated FDA recalls Z-2848-2020
- Recall initiated FDA recalls Z-2888-2020
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- OWB
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent