ClotTriever XL Catheter
510(k) K223210
Active
510(k)
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK223210 | Date27 Apr 2023 | StatusActive |
Recalls
-
- Ongoing Z-2587-2024
Thrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- QEW
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.5150
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent