Coaguchek Pst System

by Boehringer Mannheim Corp. 510(k)

Active United States

510(k) K962571

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK962571 Date22 Apr 1997ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

4 recalls · FDA

    • Terminated Z-0737-2007 FDA Unclassified (pre-2012) initiated 28 Feb 2007

    The product was erroneously distributed to home users in contrast to its use for healthcare professionals only.

    • Terminated Z-0227-2007 FDA Unclassified (pre-2012) initiated 18 Oct 2006

    Erroneous Test Results: Monitor may display 'error' message or report falsely elevated patient results caused by insufficient amounts of active ingredient (thromboplastin) in the test strips.

    • Terminated Z-0098-06 FDA Unclassified (pre-2012) initiated 28 Mar 2005

    Breaking the glass ampoule without first taking precautions may result in cuts/user injury from the glass.

    • Terminated Z-1293-03 FDA Unclassified (pre-2012) initiated 12 Sep 2003

    Erroneous patient results: may be falsely elevated or lowered, due to a packaging defect that allows air and moisture to enter the sealed pouch.

Specification

510(k)

510(k) Number
FDA Product Code
GJS
Device Class
Class II
Regulation
21 CFR Part 864.7750
Advisory Committee
Hematology
FDA Decision
SESE Substantially Equivalent