Coaguchek System, Coaguchek System for Prothrombin Time Self Testing
by Boehringer Mannheim Corp. 510(k)
Active
510(k) K974569
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK974569 | Date9 Jan 1998 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0737-2007
The product was erroneously distributed to home users in contrast to its use for healthcare professionals only.
-
- Terminated Z-0227-2007
Erroneous Test Results: Monitor may display 'error' message or report falsely elevated patient results caused by insufficient amounts of active ingredient (thromboplastin) in the test strips.
-
- Terminated Z-0099-06
Breaking the glass ampoule without first taking precautions may result in cuts/user injury from the glass.
-
- Terminated Z-1293-03
Erroneous patient results: may be falsely elevated or lowered, due to a packaging defect that allows air and moisture to enter the sealed pouch.
-
- Terminated Z-0720-03
Discrepancy between INR and the % Quick displayed on the meter.
Specification
- 510(k) Number
- FDA Product Code
- JPA
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.5425
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent