CoaguChek XS System
by Roche Diagnostics GmbH Germany EUDAMED510(k)
Ref 04625412170 EMDN W02020602
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDD EUDAMED | NumberB-04015630945719 | Date24 Aug 2022 | StatusActive |
| United States | FDA 510(k) 510(k) | NumberK060978 | Date11 Aug 2006 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- United States
- European Union
Recalls
-
- Terminated Z-0803-2021
Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
-
- Terminated Z-1584-2013
Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
-
- Terminated Z-2002-2009
The recommended cleaning procedures for the meter may cause a malfunction and falsely elevated test result reports.
Specification
- Model / Reference
- 04625412170
- Primary DI / UDI-DI
- 04015630945719
- Basic UDI-DI
- B-04015630945719
- EMDN Code
- W02020602 Rapid test hemostasis instruments
- Device Class
- Class IVD devices for self-testing
- Market of Registration
- Germany
Manufacturer details
- EUDAMED SRN
- DE-MF-000006260
Other devices from Roche Diagnostics GmbH
Properties & raw data
Show properties, registry sources and raw records
Other properties (7)
- Sterile: No
- Human Tissues: No
- Animal Tissues: No
- Microbial Substances: No
- Reagent: No
- Professional Testing: No
- Companion Diagnostics: No
Sources (2)
- EUDAMED 83f616e0-d5c3-4112-b950-8d5760d2081b
- FDA 510(k) K060978