# Cobas 6000 Series System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/cobas-6000-series-system-cem>
- Manufacturer: [Roche Diagnostics](https://devicerts.com/manufacturers/roche-diagnostics-corp) (United States, FDA 510(k) applicant)
- Name in the registry: COBAS 6000 SERIES SYSTEM
- Device class: Class II (US FDA)
- FDA product code: CEM

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K060373 · decision 2006-03-13 · decision SESE · status: Active

## Recalls

- Z-2476-2019 · Class II · terminated · initiated 2019-07-18 · matched through shared submission K060373, may concern other models: Quality issue with high pressure solenoid valves may cause inaccurate results.
- Z-1268-2018 · Class II · terminated · initiated 2017-12-18 · matched through shared submission K060373, may concern other models: Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 sample tubes without tube adapters.
- Z-0909-2018 · Class II · terminated · initiated 2017-07-18 · matched through shared submission K060373, may concern other models: A warning is being added to the cobas 6000 MODULAR Series System - c501 and the cobas 8000 MODULAR Analyzer Series - c502 Operator Manuals to prevent operator injury during maintenance.
- Z-0378-2018 · Class II · terminated · initiated 2017-06-29 · matched through shared submission K060373, may concern other models: Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.
- Z-2049-2017 · Class II · terminated · initiated 2017-03-21 · matched through shared submission K060373, may concern other models: Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the affected sample probe may dip into the sample material deeper than intended. This may lead to carryover and medical risk cannot be excluded.
- Z-0096-2017 · Class II · terminated · initiated 2016-07-15 · matched through shared submission K060373, may concern other models: Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
- Z-1791-2014 · Class II · terminated · initiated 2014-05-21 · matched through shared submission K060373, may concern other models: Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
- Z-1792-2014 · Class II · terminated · initiated 2014-05-21 · matched through shared submission K060373, may concern other models: Under certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
- Z-0542-2009 ·  · terminated · initiated 2008-09-15 · matched through shared submission K060373, may concern other models: Test results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
- Z-0576-2008 ·  · terminated · initiated 2007-12-04 · matched through shared submission K060373, may concern other models: Incorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported.

## Identifiers

- 510(k) Number: K060373 [FDA 510(k)]
- FDA product code: CEM [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 862.1600 [FDA 510(k)]
- Advisory committee: Clinical Chemistry [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Cobas 6000 Series System, Roche Diagnostics. deviCERTS medical device catalogue. https://devicerts.com/devices/cobas-6000-series-system-cem (data as of 2026-11-24).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-24
- JSON: <https://devicerts.com/devices/cobas-6000-series-system-cem.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/cobas-6000-series-system-cem
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.