Cobas Integra Reagent Cassettes & Ancillary Reagents
by Roche Diagnostic Systems, Inc.
510(k) K963292
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK963292 | Date31 Oct 1996 | StatusActive |
Recalls
-
- Terminated Z-1824-2011
Internal investigations have confirmed imprecision and erroneous patient results (up to 35% discrepant results) for specific lots of the Calcium assay. Internal investigations revealed the presence of precipitate in the R2 of the affected calcium reagent lots. The precipitate observed was classified as o-cresolphthalein complexone (o-CPC), a reactive component of the calcium R2. The root cau
-
- Terminated Z-1823-2011
The firm has recieved complaints of imprecision and erroneous patient results for calcium on the COBAS INTEGRA and COBAS c 501 system using reagent lots 62601901 and 63020401.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CIC
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1145
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 25 Sep 2026
The Cobas Integra Reagent Cassettes & Ancillary Reagents are in vitro diagnostic devices used for automated testing on the Cobas Integra system. They are designed for clinical chemistry assays and utilize absorbance, fluorescence polarization, and ion-selective electrode methodologies.
These reagent cassettes are barcode-identified, preparation-free, and stored on-board at 2-8°C. They are supplied for use in clinical laboratory settings and are authorized under FDA 510(k) clearance K963292.
Other devices from Roche Diagnostic Systems, Inc.
Questions
AI-generated answers, 25 Sep 2026
What is the Cobas Integra Reagent Cassettes & Ancillary Reagents used for?
These reagent cassettes are in vitro diagnostic devices designed for automated clinical chemistry assays on the Cobas Integra system, utilizing absorbance, fluorescence polarization, and ion-selective electrode methodologies to perform laboratory tests on patient samples.
How are the Cobas Integra Reagent Cassettes supplied and stored?
The reagent cassettes are supplied as barcode-identified, preparation-free units intended for on-board storage at 2-8°C within the Cobas Integra system, ensuring stability and readiness for use in clinical laboratory settings.
Are the Cobas Integra Reagent Cassettes single-use or reusable?
The reagent cassettes are single-use devices; each cassette is designed for one-time use in the automated analyzer and is discarded after the assay is completed, as they contain pre-measured reagents that cannot be replenished or reused.