Skip to content

Cobe Smarxt Optimin Surface Modified Hollow Fiber Membrane Oxygenator

by Cobe Cardiovascular, Inc.

510(k) K013480

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK013480 Date16 Nov 2001ClearedStatusActive

Recalls

2 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-1424-06 FDA Unclassified (pre-2012) initiated 26 Jun 2006
    • Terminated Z-1427-06 FDA Unclassified (pre-2012) initiated 26 Jun 2006

    Cooling and rewarming time is extended. Efficiency of the heat exchangers is reduced because many of the folds that are critical to heat exchange on the blood side of the units were blocked with a metalic material, thus decreasing the surface area needed for proper heat transfer.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
DTZ
Device Class
Class II
Regulation
21 CFR Part 870.4350
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Other devices from Cobe Cardiovascular, Inc.