# Codman Ap-04009

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/codman-ap04009>
- Manufacturer: [Cerenovus, Inc.](https://devicerts.com/manufacturers/cerenovus-inc) (United States)
- Name in the registry: CODMAN AP-04009
- Device class: Class III (US FDA)
- FDA product code: LKK
- GMDN: Implantable intrathecal infusion pump, nonprogrammable (34071)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 10886704072881 · listed 2014-09-20 · status: Listed (GUDID: not in commercial distribution)
- **GUDID** · FDA PMA · United States · P890055 · decision 1996-03-11 · decision APPR · status: Active · from GUDID

## Recalls

- Z-2580-2026 · Class II · completed · initiated 2026-05-11: There is the potential for leakage along the drug pathway from the pump through the end of the catheter.
- Z-1522-2022 · Class I · terminated · initiated 2022-07-10: Higher than expected flow rate.
- Z-1571-2014 · Class II · terminated · initiated 2015-11-10: A small number of specific pumps may trigger a low battery alarm at the beginning of the implanted life.
- Z-1572-2014 · Class II · terminated · initiated 2015-11-10: A small number of specific pumps may trigger a low battery alarm at the beginning of the implanted life.
- Z-1684-2015 · Class II · terminated · initiated 2015-03-05: Needle is leaking from hub/needle cannula connection
- Z-2293-2013 · Class I · terminated · initiated 2013-08-13: Drug flow rates exceeding programmed flow rates.
- Z-2294-2013 · Class I · terminated · initiated 2013-08-13: Drug flow rates exceeding programmed flow rates.
- Z-2295-2013 · Class I · terminated · initiated 2013-08-13: Drug flow rates exceeding programmed flow rates.
- Z-2296-2013 · Class I · terminated · initiated 2013-08-13: Drug flow rates exceeding programmed flow rates.
- Z-2297-2013 · Class I · terminated · initiated 2013-08-13: Drug flow rates exceeding programmed flow rates.
- Z-2298-2013 · Class I · terminated · initiated 2013-08-13: Drug flow rates exceeding programmed flow rates.
- Z-2299-2013 · Class I · terminated · initiated 2013-08-13: Drug flow rates exceeding programmed flow rates.
- Z-2300-2013 · Class I · terminated · initiated 2013-08-13: Drug flow rates exceeding programmed flow rates.
- Z-1690-2013 · Class I · terminated · initiated 2013-06-07: Miscalibrated Fill Level Sensor may affect dosing level
- Z-1691-2013 · Class I · terminated · initiated 2013-06-07: Miscalibrated Fill Level Sensor may affect dosing level

## Identity

- Trade name: CODMAN [FDA GUDID (AccessGUDID)]
- Generic name: Implantable intrathecal infusion pump, nonprogrammable [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: AP-04009 [FDA GUDID (AccessGUDID)]; AP04009 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 10886704072881 [FDA GUDID (AccessGUDID)]
- PMA number: P890055 [FDA GUDID (AccessGUDID)]
- FDA product code: LKK [FDA GUDID (AccessGUDID)]
- GMDN term: Implantable intrathecal infusion pump, nonprogrammable (34071) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class III [FDA GUDID (AccessGUDID)]
- GUDID distribution status: Not in Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Supplied sterile: Yes [FDA GUDID (AccessGUDID)]
- Single use: Yes [FDA GUDID (AccessGUDID)]
- Implantable: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Expiration date: Yes [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Codman Ap-04009, Cerenovus, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/codman-ap04009 (data as of 2026-11-23).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/codman-ap04009.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/codman-ap04009
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.