Compass

by Scanditronix Wellhofer GmbH

510(k) K072374

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK072374 Date7 Dec 2007ClearedStatusActive

Recalls

3 recalls · FDA

    • Terminated Z-1212-2015 FDA Class II initiated 2 Feb 2015

    Error in the software. During internal tests of the current development version of the Compass SW it was found that dose reconstruction for DMLC plans when the jaws move or mlc leaves reverse during beam on will not be scaled correctly for all control points.

    • Terminated Z-0168-2015 FDA Class II initiated 25 Sep 2014

    Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.

    • Terminated Z-0456-2014 FDA Class II initiated 20 Aug 2013

    The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IYE
Device Class
Class II
Regulation
21 CFR Part 892.5050
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent