  "format": "devicerts.device.v1",
  "cite_as": "Comprehensive Reverse Shoulder, Biomet, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/comprehensive-reverse-shoulder-phx-k080642 (data as of 2026-11-21).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/comprehensive-reverse-shoulder-phx-k080642",
    "json_url": "https://devicerts.com/devices/comprehensive-reverse-shoulder-phx-k080642.json",
    "markdown_url": "https://devicerts.com/devices/comprehensive-reverse-shoulder-phx-k080642.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-21",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/comprehensive-reverse-shoulder-phx-k080642",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Comprehensive Reverse Shoulder",
    "registry_name": "COMPREHENSIVE REVERSE SHOULDER",
    "url": "https://devicerts.com/devices/comprehensive-reverse-shoulder-phx-k080642",
    "manufacturer": {
      "name": "Biomet, Inc.",
      "url": "https://devicerts.com/manufacturers/biomet-inc",
      "country": "United States",
      "country_basis": "FDA 510(k) applicant",
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K080642",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "PHX"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K080642",
      "number_label": null,
      "device_udi_di": null,
      "date": "2008-07-09",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2144-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-04-17",
      "reason": "Potential presence of elevated endotoxin levels that exceed the specification limit",
      "applies": "this device",
      "via": "K080642"
    },
    {
      "number": "Z-1750-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-03-13",
      "reason": "There were unknown bioburden levels prior to sterilization, which may affect sterility of the product.",
      "applies": "this device",
      "via": "K080642"
    },
    {
      "number": "Z-1859-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-02-17",
      "reason": "Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-1103-2017",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2016-12-20",
      "reason": "Higher than anticipated rate of fracturing due to design.  Risk of revision surgery due to fracturing.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-1548-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-05-15",
      "reason": "Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction.\nWhen the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2545-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2541-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2549-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2548-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2547-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2550-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2546-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2542-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2544-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-2543-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-22",
      "reason": "The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-0712-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2010-09-13",
      "reason": "The firm received complaints regarding fracturing of the device.",
      "applies": "shared submission",
      "via": "K080642"
    },
    {
      "number": "Z-0713-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2010-09-13",
      "reason": "The firm received complaints regarding fracturing of the device.",
      "applies": "shared submission",
      "via": "K080642"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K080642",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "PHX",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 888.3660",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Orthopedic",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}