Skip to content

Comprehensive® Shoulder System 118001

by Biomet Orthopedics, Inc.

Listed FDA GUDID listing, no authorization on record

Valid until

Not applicable

Database listing, no authorization

FDA recalls

1 ongoing

This device is named in the recall

Documents

None

No authorization documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source GUDID, synced 23 Oct 2026. What this means

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00880304217249 UDI-DI Date24 Oct 2015ListedStatusListed GUDID: in commercial distribution

Recalls

18 recalls · 1 ongoing · FDA

    • Ongoing Z-1403-2025 FDA Class II initiated 3 Mar 2025

    Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.

    • Terminated Z-2144-2020 FDA Class II initiated 17 Apr 2020

    Potential presence of elevated endotoxin levels that exceed the specification limit

    • Terminated Z-1750-2020 FDA Class II initiated 13 Mar 2020

    There were unknown bioburden levels prior to sterilization, which may affect sterility of the product.

    • Terminated Z-1859-2017 FDA Class II initiated 17 Feb 2017

    Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.

    • Terminated Z-1103-2017 FDA Class I initiated 20 Dec 2016 on shared 510(k) K080642

    Higher than anticipated rate of fracturing due to design. Risk of revision surgery due to fracturing.

    • Terminated Z-1548-2013 FDA Class II initiated 15 May 2013 on shared 510(k) K080642

    Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.

  • 10 recalls, same reason

    • Terminated Z-2541-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2549-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2548-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2547-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2550-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2546-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2542-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2545-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2544-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011
    • Terminated Z-2543-2011 FDA Unclassified (pre-2012) initiated 22 Apr 2011

    The firm is initiating this recall following an investigation which identified the rare possibility that certain units of Comprehensive Reverse Shoulder Humeral Trays may contain a locking ring that is incorrectly assembled.

  • 2 recalls, same reason

    • Terminated Z-0712-2011 FDA Unclassified (pre-2012) initiated 13 Sep 2010
    • Terminated Z-0713-2011 FDA Unclassified (pre-2012) initiated 13 Sep 2010

    The firm received complaints regarding fracturing of the device.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — GUDID listing Listed

Specification

GUDID

Trade Name
Comprehensive® Shoulder System
Generic Name
Stem-fixed metallic humeral head prosthesis
Model / Ref / Catalog
118001
Primary DI / UDI-DI
00880304217249
FDA Product Code
MBF, KWS, KWT, HSD, PHX
GMDN Term
Stem-fixed metallic humeral head prosthesis
Device Class
Class II
Regulation Number
888.3670, 888.3660, 888.3650, 888.3690
GUDID distribution status
In Commercial Distribution

Other devices from Biomet Orthopedics, Inc.

See all devices from Biomet Orthopedics, Inc. →

Properties

  • Sterile
  • Single Use
  • Implantable
  • Prescription Use
  • Lot / Batch Number
  • Expiration Date

Other properties (8)

  • Contains Latex: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Serial Number: No