# Computed Tomography X-ray System

> Active: European Union and United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/computed-tomography-x-ray-system-88000870>
- Manufacturer: [Shanghai United Imaging Healthcare Co., Ltd.](https://devicerts.com/manufacturers/shanghai-united-imaging-healthcare-co-ltd) (China)
- Device class: Class IIb (EU MDD / EU MDR)
- EMDN: Z11030607 COMPUTED TOMOGRAPHS - 256 LAYERS OR MORE

## Registrations and authorizations

- **EUDAMED** · CE/MDD · European Union · Basic UDI-DI 697157683109NB · UDI-DI 06971576831098 · registered 2025-04-08 · status: Active
- **EUDAMED** · CE/MDR · European Union · Basic UDI-DI 697157683109NB · UDI-DI 06971576831098 · registered 2025-04-08 · status: Active
- **510(k)** · FDA 510(k) · United States · K172135 · issued 2018-03-02 · status: Active

## Recalls

- Z-3279-2026 · Class II · ongoing · initiated 2026-08-11 · matched through shared submission K172135, may concern other models: Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
- Z-3280-2026 · Class II · ongoing · initiated 2026-08-11 · matched through shared submission K172135, may concern other models: Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
- Z-0219-2026 · Class II · ongoing · initiated 2025-09-12 · matched through shared submission K172135, may concern other models: Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Z-0220-2026 · Class II · ongoing · initiated 2025-09-12 · matched through shared submission K172135, may concern other models: Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Z-0447-2021 · Class II · ongoing · initiated 2020-08-05 · matched through shared submission K172135, may concern other models: A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.
- Z-2703-2020 · Class II · ongoing · initiated 2020-06-20 · matched through shared submission K172135, may concern other models: The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge.
- Z-2522-2020 · Class II · terminated · initiated 2020-05-08 · matched through shared submission K172135, may concern other models: Two issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift. This may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

## Identity

- Device name: Computed Tomography X-ray System [EUDAMED]
- Model / reference: uCT 960+ [EUDAMED]; 88000870 [EUDAMED]; 88000228 [EUDAMED]

## Identifiers

- Primary DI / UDI-DI: 06971576831098 [EUDAMED]
- Basic UDI-DI: 697157683109NB [EUDAMED]; B-06971576831098 [EUDAMED]
- EMDN code: Z11030607 [EUDAMED]

## Classification

- Device class: Class IIb [EUDAMED]
- System or procedure pack: System which is a device in itself [EUDAMED]
- EU registration type: Standard device [EUDAMED]; Legacy device (under a directive) [EUDAMED]
- EU market of registration: Poland [EUDAMED]

## Properties

- Measuring function: Yes [EUDAMED]
- Supplied sterile: No [EUDAMED]
- Sterilise before use: No [EUDAMED]
- Implantable: No [EUDAMED]
- Reusable: No [EUDAMED]
- Contains latex: No [EUDAMED]
- Reprocessed single-use device: No [EUDAMED]
- Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction): No [EUDAMED]
- Incorporates a substance derived from human blood or plasma: No [EUDAMED]
- Contains human tissues or cells: No [EUDAMED]
- Contains animal tissues or cells: No [EUDAMED]
- Contains endocrine-disrupting substances: No [EUDAMED]
- Administering medicine: No [EUDAMED]
- Incorporates a medicinal substance: No [EUDAMED]

## Source and citation

- Cite as: Computed Tomography X-ray System, Shanghai United Imaging Healthcare Co., Ltd. deviCERTS medical device catalogue. https://devicerts.com/devices/computed-tomography-x-ray-system-88000870 (data as of 2026-11-29).
- Data sources: [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-24
- Statuses computed as of: 2026-11-29
- JSON: <https://devicerts.com/devices/computed-tomography-x-ray-system-88000870.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/computed-tomography-x-ray-system-88000870
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.