Skip to content

Conmed: Stat 2 and Stat 2 Pumpette Extension Sets and Masterflow and Masterflow Pumpette Primary Sets

by Conmedcorp

510(k) K964844

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

12 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK964844 Date24 Mar 1997ClearedStatusActive

Recalls

12 recalls · FDA

    • Terminated Z-0805-2018 FDA Class II initiated 19 Jun 2017

    Intravascular (I.V.) fluid administration sets are exhibiting inaccurate or inconsistent flow rates.

  • 5 recalls, same reason

    • Terminated Z-0684-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0685-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0688-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0687-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0686-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011

    ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the body of the compensating controller. The investigation of the complaints determined that the STAT 2 Pumpette Compensating Controller on several complaint samples had been assembled incorrectly.

  • 6 recalls, same reason

    • Terminated Z-0636-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0632-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0631-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0633-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0630-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011
    • Terminated Z-0635-2012 FDA Unclassified (pre-2012) initiated 15 Apr 2011

    ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined that certain units were assembled with an incorrect component which could result in the devices exhibiting a different flow rate and might reduce the ability of the user to regulate the device at low flow rates.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
FPA
Device Class
Class II
Regulation
21 CFR Part 880.5440
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Other devices from Conmedcorp