Connector System
510(k) K172194
Connector System is a Class II device made by Orthofix, Inc., cleared under FDA 510(k) K172194 on 18 Aug 2017, active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK172194 | Date18 Aug 2017 | StatusActive |
Official records
- FDA 510(k) database K172194
Recalls
-
- Terminated Z-1363-2018
- Terminated Z-1364-2018
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Timeline
- Cleared FDA 510(k) K172194
- Recall initiated FDA recalls Z-1363-2018
- Recall initiated FDA recalls Z-1364-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- NKG
- Device class
- Class II
- Regulation
- 21 CFR Part 888.3075
- Advisory committee
- Orthopedic
- FDA decision
- SESE Substantially Equivalent