Skip to content

Contak Renewal TR Pulse Generator

by Guidant Corp.

PMA P030005

Active United States

PMA

Valid until

No expiry

PMA approval is not time-limited

FDA recalls

17 ongoing

This device is named in the recall

Documents

PMA P030005↗

FDA record

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA PMA, synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA PMAAPPR Approved NumberP030005 Date26 Jan 2004ApprovedStatusActive

Recalls

20 recalls · 17 ongoing · FDA

    • Ongoing Z-1771-2026 FDA Class I initiated 19 Mar 2026

    Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia

  • 5 recalls, same reason

    • Ongoing Z-0104-2026 FDA Class I initiated 20 Aug 2025
    • Ongoing Z-0101-2026 FDA Class I initiated 20 Aug 2025
    • Ongoing Z-0103-2026 FDA Class I initiated 20 Aug 2025
    • Ongoing Z-0105-2026 FDA Class I initiated 20 Aug 2025
    • Ongoing Z-0102-2026 FDA Class I initiated 20 Aug 2025

    Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

  • 2 recalls, same reason

    • Ongoing Z-0827-2025 FDA Class I initiated 12 Dec 2024
    • Ongoing Z-0826-2025 FDA Class I initiated 12 Dec 2024

    A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

  • 3 recalls, same reason

    • Ongoing Z-2019-2021 FDA Class I initiated 3 Jun 2021
    • Ongoing Z-2020-2021 FDA Class I initiated 3 Jun 2021
    • Ongoing Z-2018-2021 FDA Class I initiated 3 Jun 2021

    The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population

  • 2 recalls, same reason

    • Ongoing Z-2002-2021 FDA Class II initiated 3 Jun 2021
    • Ongoing Z-2003-2021 FDA Class II initiated 3 Jun 2021

    Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.

  • 4 recalls, same reason

    • Ongoing Z-0365-2018 FDA Class II initiated 7 Dec 2017
    • Ongoing Z-0360-2018 FDA Class II initiated 7 Dec 2017
    • Ongoing Z-0366-2018 FDA Class II initiated 7 Dec 2017
    • Ongoing Z-0361-2018 FDA Class II initiated 7 Dec 2017

    Boston Scientific has received reports of intermittent over-sensing of the Minute Ventilation (MV) sensor signal with certain Boston Scientific pacemaker and cardiac resynchronization therapy pacemaker systems (pacemakers).

  • 2 recalls, same reason

    • Terminated Z-3258-2018 FDA Class II initiated 10 Sep 2018 on shared PMA P030005
    • Terminated Z-3259-2018 FDA Class II initiated 10 Sep 2018 on shared PMA P030005

    Boston Scientific has identified a subset of active pacemaker and CRT-Ps systems with an elevated potential for early pacemaker replacement due to hydrogen-induced accelerated battery depletion.

    • Terminated Z-1298-06 FDA Unclassified (pre-2012) initiated 24 Jun 2006

    Guidant has determined that low-voltage capacitors from a single component supplier may malfunction in a manner that can lead to premature battery depletion without warning in the affected devices.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA PMA Active

Specification

PMA

Full name
Contak Renewal Tr
Trade Name
Contak Renewal Models H125 and H120 with the Model 2865 Version 1.8 Application Software
Generic Name
Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)
PMA Number
FDA Product Code
NKE
Device Class
Class III
Advisory Committee
Cardiovascular
FDA Decision
APPR Approved

Overview

AI-generated summary, 18 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The Contak Renewal TR is an implantable cardiac resynchronization therapy-pacemaker (CRT-P), classified as a pulse generator designed to manage moderate to severe heart failure. It provides atrial-ventricular tracking modes to preserve atrioventricular synchrony and adaptive-rate pacing to adjust pacing rates in response to physical activity, supporting patients with left ventricular dysfunction and prolonged QRS duration.

This device is supplied sterile and intended for single-use implantation, with regulatory approval under an FDA Premarket Approval (PMA). It is intended for use in clinical settings by trained healthcare professionals and requires specific software (Model 2865, Version 1.8) for programming. The device is MRI-unsafe and must be stored according to manufacturer instructions to maintain sterility and functionality.

Other devices from Guidant Corp.

See all devices from Guidant Corp. →