Corded 2D Barcode Imager
by Instrumentation Laboratory Company United States EUDAMEDGUDID
Ref 00024015859 Product code CHLEMDN W0201040180
Regulatory status
Recall ongoing. 26 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/IVDR EUDAMED | Number8426950869634BJ | Date15 Apr 2024 | StatusActive |
| United States | GUDID listing GUDID | Number08426950869634 | Date29 Mar 2017 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- European Union
- United States
Recalls
-
- Ongoing Z-2900-2026
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
-
- Ongoing Z-2823-2026
- Ongoing Z-2825-2026
- Ongoing Z-2824-2026
- Ongoing Z-2826-2026
- Ongoing Z-2821-2026
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
-
- Ongoing Z-1529-2026
- Ongoing Z-1533-2026
- Ongoing Z-1524-2026
- Ongoing Z-1530-2026
- Ongoing Z-1541-2026
- Ongoing Z-1536-2026
- Ongoing Z-1528-2026
- Ongoing Z-1526-2026
- Ongoing Z-1525-2026
- Ongoing Z-1542-2026
- Ongoing Z-1535-2026
- Ongoing Z-1531-2026
- Ongoing Z-1540-2026
- Ongoing Z-1543-2026
- Ongoing Z-1532-2026
- Ongoing Z-1539-2026
- Ongoing Z-1534-2026
- Ongoing Z-1538-2026
- Ongoing Z-1527-2026
- Ongoing Z-1537-2026
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
Specification
- Trade Name
- Corded 2D Barcode Imager EUDAMEDWand, Bar Code GUDID
- Generic Name
- Barcode reader
- Model / Ref / Catalog
- 00024015859
- Primary DI / UDI-DI
- 08426950869634
- Basic UDI-DI
- 8426950869634BJ
- Direct Marketing DI
- 08426950869634
- FDA Product Code
- CHL
- EMDN Code
- W0201040180 Blood gas analysers - hardware accessories
- GMDN Term
- Barcode reader
- Device Class
- Class A EUDAMEDClass II GUDID
- Regulation Number
- 862.1120
- Market of Registration
- Italy
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class A | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000035586
- Authorised representative
- Instrumentation Laboratory SpA IT-AR-000008169
Other devices from Instrumentation Laboratory Company
Properties & raw data
Show properties, registry sources and raw records
Other properties (20)
- Sterile: No
- Single Use: No
- Implantable: No
- Contains Latex: No
- Human Tissues: No
- Animal Tissues: No
- Microbial Substances: No
- Reagent: No
- Self Testing: No
- Near-Patient Testing: No
- Companion Diagnostics: No
- Surgical Instrument: No
- New Device: No
- Over-the-Counter: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (2)
- EUDAMED a9f0bf2b-e578-45f5-b4a7-a1c20e21df83
- FDA UDI c9d075d9-b43a-4633-add7-7574bacd8a80