Cortera 150150-85095
by Surgalign Spine Technologies
Cortera 150150-85095 is a bone-screw internal spinal fixation system (FDA Class II; FDA product code NKB, Thoracolumbosacral pedicle screw system) made by Surgalign Spine Technologies, cleared under FDA 510(k) K221403 on 19 Aug 2022 and listed in FDA GUDID since 6 Jan 2023 (UDI-DI 00840311201587), active as of 23 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00840311201587 | Date6 Jan 2023 | StatusListed |
| United States | FDA 510(k)SESE | NumberK221403 | Date19 Aug 2022 | StatusActive |
Official records
- AccessGUDID record 00840311201587
- FDA 510(k) database K221403
From the FDA record
- Device description
- Screw, Poly, MIS, Cann, 8.5x95
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K221403 decided 19 Aug 2022 Active
- FDA GUDID UDI-DI 00840311201587 listed 6 Jan 2023 Listed
Recalls
-
- Ongoing Z-0125-2025
Spinal fixation system implant screws marked "DNI" are not intended to be implanted and may not have been manufactured in accordance with released manufacturing specifications, and if implanted the screws may not have adequate strength or performance.
-
- Ongoing Z-0846-2024
Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.
Timeline
- Cleared FDA 510(k) K221403
- Listed FDA GUDID 00840311201587
- Recall initiated FDA recalls Z-0846-2024
- Recall initiated FDA recalls Z-0125-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Cortera
- Generic name
- Bone-screw internal spinal fixation system, non-sterile
- Model / reference
- 150150-85095
- Primary DI / UDI-DI
- 00840311201587
- 510(k) Number
- FDA product code
- NKB
- GMDN term
- Bone-screw internal spinal fixation system, non-sterile
- Device class
- Class II
- Regulation number
- 888.3070
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
Similar devices
Other devices filed under FDA product code NKB (Thoracolumbosacral pedicle screw system).
- Erisma™-LP 18455514 by Clariance, Inc.
- Overwatch Spine System by Amendia, Inc.
- SYNTHES MATRIX SYSTEM by Synthes (Usa)
- Daytona® Small Stature Spinal System by SeaSpine Orthopedics Corporation
- Seneka II Polyscrew Pedicle Fixation System by Senecka Spine
- Erisma™-LP 18415535 by Clariance, Inc.
- DEL MAR MONOAXIAL PEDICLE SCREW SYSTEM by Allez Spine, LLC
- Momentum® Posterior Spinal Fixation System by ULRICH MEDICAL USA, INC.
- Zavation Connector System by Zavation Medical Products, LLC
- Ibis® Pedicle Screw System by ORTHO DEVELOPMENT CORP.