Coulter Lh 750 Hematology Analyzer, Model 6605632
by Beckman Coulter, Inc. 2 variants
510(k) K011342
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK011342 | Date9 Jul 2001 | StatusActive |
Recalls
-
- Terminated Z-3113-2017
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
-
- Terminated Z-2480-2015
- Terminated Z-2481-2015
Beckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.
-
- Terminated Z-0470-2014
Beckman Coulter is recalling the LH 750 and LH 780 Hematology Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), white blood cell differential and/or reticulocytes.
-
- Terminated Z-0113-2014
The check valves may fail on the Coulter LH 750 Hematology Analyzer (PN 6605632 and A85570) and Coulter LH 780 Hematology Analyzer (PN 723585 and A90728). A failure may result in an air or liquid leak at the connection to the Vacuum Overflow Tank waste line or at the Backwash Tank drain line.
-
- Terminated Z-1242-2012
The recall was initiated because Beckman Coulter has confirmed that the Auto Stop and Auto Transmit features for Quality Control become disabled after scanning assay values from Coulter 5C and Retic-C Cell Controls assay sheets using the 2D Barcode scanner.
-
- Terminated Z-1953-2011
This recall was initiated because Beckman Coulter identified a problem with some date fields that are transmitted to the Laboratory Information System (LIS). Starting Jan 1, 2011, the date fields will transmit as a blank field to the LIS if the year is greater than 2010. The following dates are affected: Run date, Control expiration date, Date of birth (For individuals born in 2011 or later), a
-
- Terminated Z-0312-2008
Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2)- When manually entering a Patient ID, if a blank space is entered between th
-
- Terminated Z-0566-2007
There is a potential for erroneous yet credible results for Body Fluids on the LH700 series hematology analyzers when a cassette label fails to read and the subsequent sample is cycled in Body Fluid mode.
-
- Terminated Z-0474-06
There is a risk of sample misidentification when processing samples in the Manual aspiration mode.
-
- Terminated Z-0905-04
If a customer enters a dilution factor for Sample B prior to the results from Sample A being transfered to the workstation, the predilute multiplication factor will be applied erroneously to the Sample A's results once they are received at the workstation.
-
- Terminated Z-0541-03
Biohazard situation exists for users.
-
- Terminated Z-0669-03
Incorrect Hemoglobin result can be reported with software version 2A and higher.
-
- Terminated Z-0571-03
Device design allows for cross contamination issue to occur, possibly affecting assays and results reports.
-
- Terminated Z-0415-03
Software vote out logic changed. Erroneous white blood cell counts.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- GKZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.5220
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent