Skip to content

Cr-30

by Nakanishi Inc.

Ref 10000977Model MDA0000030

Active European Union. Also listed in GUDID (US).

CE MDR

Valid until

Not listed

EUDAMED holds no validity date

FDA recalls

None ongoing

1 closed

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source EUDAMED, GUDID, synced 25 Oct 2026. What this means

Regulatory status

as of registry sync 25 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number45602249800002177C Basic UDI-DI 04589551389001UDI-DIDate31 Jul 2026RegisteredStatusActive
United StatesGUDID listing GUDID Number04589551389001 UDI-DI Date20 Apr 2026ListedStatusListed GUDID: in commercial distribution

Certificates

0 certificates · EUDAMED

EUDAMED lists no certificate for this device's Basic UDI-DI.

A certificate issued for a device group may exist without being linked here.

Recalls

1 recall · FDA

    • Terminated Z-2673-2016 FDA Class II initiated 8 Jul 2016

    Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States GUDID listing Listed

Specification

EUDAMED · GUDID

Generic Name
Ultrasonic surgical system torque wrench, reusable
Model / Reference
MDA0000030 EUDAMED
10000977 EUDAMED
CR-30 GUDID
Primary DI / UDI-DI
04589551389001
Basic UDI-DI
45602249800002177C
510(k) Number
FDA Product Code
DZI, ELC
EMDN Code
Z12119080 Various dental stomatology instruments - hardware accessories
GMDN Term
Ultrasonic surgical system torque wrench, reusable
Device Class
Class I EUDAMED
Class II GUDID
Regulation Number
872.4120, 872.4850
EU market of registration
Germany
GUDID distribution status
In Commercial Distribution
Sterilization Method
Moist Heat or Steam Sterilization

Truth per registry

2 sources · 2 fields differ

FieldEUDAMEDGUDID
Model / ReferenceEUDAMED
MDA0000030
10000977
GUDID
CR-30
Device ClassEUDAMED
Class I
GUDID
Class II

Other devices from Nakanishi Inc.

See all devices from Nakanishi Inc. →

Properties

  • Prescription Use
  • Lot / Batch Number
  • Manufacturing Date

Other properties (21)

  • Sterile: No
  • Single Use: No
  • Implantable: No
  • Reusable: No
  • Contains Latex: No
  • Reprocessed: No
  • CMR Substance: No
  • Human-derived: No
  • Human Tissues: No
  • Animal Tissues: No
  • Endocrine Disruptor: No
  • Administering Medicine: No
  • Measuring Function: No
  • Incorporates Medicinal Product: No
  • Over-the-Counter: No
  • Convenience Kit: No
  • Combination Product: No
  • HCT/P: No
  • Premarket Exempt: No
  • Serial Number: No
  • Expiration Date: No