Css Cannulated Screw System
by Orthohelix Surgical Designs, Inc.
510(k) K060428
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK060428 | Date7 Mar 2006 | StatusActive |
Recalls
-
- Terminated Z-0467-2011
- Terminated Z-0468-2011
- Terminated Z-0466-2011
The print for modified CSS depth gauges has inaccurate definition of the etched screw length. The error results in a length reading that is longer (5 mm in the case of the 7.0 and 4mm in the cases of the 5.5 and 4.0) than the actual length of the buried guide wire segment and even longer (10 mm in the case of the 7.0 and 8mm in the cases of the 5.5 and 4.0) than the design intention readings. This
-
- Terminated Z-1250-2010
A pinhead sized disc was found in the cannulation of a MSD-010-55-60S screw.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- HWC
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3040
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent