CUSA Clarity Ultrasonic Surgical Aspirator System
by Integra Lifesciences Corporation 7 variants
510(k) K182809
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK182809 | Date7 Dec 2018 | StatusActive |
Recalls
-
- Terminated Z-0634-2020
- Terminated Z-0633-2020
- Terminated Z-0637-2020
- Terminated Z-0614-2020
- Terminated Z-0641-2020
- Terminated Z-0653-2020
- Terminated Z-0622-2020
- Terminated Z-0643-2020
- Terminated Z-0625-2020
- Terminated Z-0620-2020
- Terminated Z-0645-2020
- Terminated Z-0631-2020
- Terminated Z-0624-2020
- Terminated Z-0636-2020
- Terminated Z-0648-2020
- Terminated Z-0650-2020
- Terminated Z-0621-2020
- Terminated Z-0619-2020
- Terminated Z-0628-2020
- Terminated Z-0651-2020
- Terminated Z-0618-2020
- Terminated Z-0647-2020
- Terminated Z-0649-2020
- Terminated Z-0616-2020
- Terminated Z-0640-2020
- Terminated Z-0644-2020
- Terminated Z-0617-2020
- Terminated Z-0646-2020
- Terminated Z-0630-2020
- Terminated Z-0629-2020
- Terminated Z-0652-2020
- Terminated Z-0615-2020
- Terminated Z-0635-2020
- Terminated Z-0632-2020
- Terminated Z-0638-2020
- Terminated Z-0639-2020
- Terminated Z-0623-2020
- Terminated Z-0627-2020
- Terminated Z-0626-2020
- Terminated Z-0642-2020
- Terminated Z-0654-2020
The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.
-
- Terminated Z-0613-2020
The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LFL
- Device Class
- U
- Advisory Committee
- Unknown
- FDA Decision
- SESE Substantially Equivalent