Cusa Selector Nxt Ultrasonic Tissue Ablation System
510(k) K081459
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK081459 | Date13 Aug 2008 | StatusActive |
Recalls
-
- Terminated Z-0491-2015
Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner.
-
- Terminated Z-2585-2011
The insulated covering of a wire cable that powers the vacuum pump in the CUSA NXT Service Module can become abraded over time resulting in the exposure of bare wire. If this bare wire were to come in contact with the metal surface of the pump, it could result in a short circuit and a loss of aspiration.
-
- Terminated Z-2638-2010
- Terminated Z-2637-2010
A ground fault malfunction at the rectifier retaining screw may cause the aspiration pump in the CUSA NXT Service module to run at maximum speed, and continue to run until the entire system is turned off. This may result in excess aspiration that requires medical intervention.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LFL
- Device Class
- U
- Advisory Committee
- Unknown
- FDA Decision
- SESE Substantially Equivalent