Cyto-Chex BCT 230353
by Streck
Cyto-Chex BCT 230353 is an evacuated blood collection tube IVD (FDA Class II; FDA product code GIM, Tubes, Vacuum Sample, With Anticoagulant) made by Streck, cleared under FDA 510(k) K080552 on 31 Jul 2008 and listed in FDA GUDID since 10 Feb 2017 (UDI-DI 40844509002206), active as of 23 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number40844509002206 | Date10 Feb 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK080552 | Date31 Jul 2008 | StatusActive |
Official records
- AccessGUDID record 40844509002206
- FDA 510(k) database K080552
From the FDA record
- Device description
- A direct-draw blood collection tube for the preservation of whole blood specimens for immunophenotyping by flow cytometry. 4ml vials.
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K080552 decided 31 Jul 2008 Active
- FDA GUDID UDI-DI 40844509002206 listed 10 Feb 2017 Listed
Recalls
-
- Ongoing Z-1684-2017
Complaints of the 2.0ml tube stopper coming loose during phlebotomy when small diameter tube adapters were not utilized. Such adapters aid with centering small diameter tubes to properly pierce the middle of the stoppers.
Timeline
- Cleared FDA 510(k) K080552
- Listed FDA GUDID 40844509002206
- Recall initiated FDA recalls Z-1684-2017
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Cyto-Chex BCT
- Generic name
- Evacuated blood collection tube IVD, K3EDTA
- Model / Ref / Catalog
- 230353
- Description
- A direct-draw blood collection tube for the preservation of whole blood specimens for immunophenotyping by flow cytometry. 4ml vials.
- Primary DI / UDI-DI
- 40844509002206
- 510(k) Number
- FDA product code
- GIM
- GMDN term
- Evacuated blood collection tube IVD, K3EDTA
- Device class
- Class II
- Regulation number
- 862.1675
- GUDID distribution status
- In Commercial Distribution
- Size
- Total Volume — 4.0 Milliliter
Similar devices
Other devices filed under FDA product code GIM (Tubes, Vacuum Sample, With Anticoagulant).
- BLOOD COLLECTION TUBE, RED by Sherwood Medical Industries
- VACUTAINER BRAND SAFETY-GARD PHLEBOTOMY SYSTEM by Bd Becton Dickinson Vacutainer Systems Preanalytic
- BD MiniDraw™ Finger Sleeve, Medium 366614 by Becton, Dickinson and Company
- PRE-DENT by Fichte, Graves, and Hunt
- GREINER VACUETTE BLOOD COLLECTION TUBE by Greiner America, Inc.
- VACUUM TUBES, STERILE (J-VAC) by Jelco Laboratories
- BD MiniDraw™ Finger Sleeve, Small 366613 by Becton, Dickinson and Company
- BACT/ALERT BLOOD COLLECTION ADAPTER CAP AND INSERT by Organon Teknika Corp.
- VACUUM BLOOD COLLECTING SYSTEM (NEO-TUBE) by Exel Intl.
- MONOJECT BLOOD COLLECTION TUBES W/SAFETY STOPPER by Sherwood Medical Co.