  "format": "devicerts.device.v1",
  "cite_as": "Cytrellis Dermal Micro-Coring System, Cytrellis Biosystems, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/cytrellis-dermal-micro-coring-system-qai (data as of 2026-11-26).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/cytrellis-dermal-micro-coring-system-qai",
    "json_url": "https://devicerts.com/devices/cytrellis-dermal-micro-coring-system-qai.json",
    "markdown_url": "https://devicerts.com/devices/cytrellis-dermal-micro-coring-system-qai.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-26",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/cytrellis-dermal-micro-coring-system-qai",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Cytrellis Dermal Micro-Coring System",
    "registry_name": "Cytrellis Dermal Micro-Coring System",
    "url": "https://devicerts.com/devices/cytrellis-dermal-micro-coring-system-qai",
    "manufacturer": {
      "name": "Cytrellis Biosystems, Inc.",
      "url": "https://devicerts.com/manufacturers/fda-fei-3020704367",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "United States"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K202517",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "QAI"
    ],
    "emdn": [],
    "gmdn": [],
    "description": {
      "text": "The Cytrellis Dermal Micro-Coring System is a powered microneedle device designed for controlled removal of micro-cores of dermal and epidermal tissue. Classified as a Class II device under FDA regulations (878.4430), it is intended for use in aesthetic procedures to address moderate to severe wrinkles in the mid and lower facial regions. The system employs precise micro-coring technology to target sagging skin without invasive surgical intervention, facilitating tissue remodeling and skin tightening.\n\nThis device is supplied as a reusable system with single-use disposable components, intended for use in clinical settings by trained practitioners. It is not MRI-compatible and requires proper sterilization and maintenance as outlined in manufacturer instructions. The system is authorized for use in the United States following FDA 510(k) clearance (K202517) under the Subcommittee for Urology (SU) and General Plastic Surgery advisory committees. Multiple registration numbers (e.g., 3007645866) correspond to its various configurations and accessories, ensuring traceability and compliance with regulatory standards.",
      "ai_generated": true,
      "generated_on": "2026-10-01",
      "note": "Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data is the record."
    },
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K202517",
      "number_label": null,
      "device_udi_di": null,
      "date": "2021-07-09",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2393-2023",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2023-06-23",
      "reason": "Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring",
      "applies": "this device",
      "via": "K202517"
    },
    {
      "number": "Z-1648-2023",
      "classification": "Class II",
      "status": "completed",
      "initiated": "2023-04-07",
      "reason": "Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.",
      "applies": "this device",
      "via": "K202517"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K202517",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "QAI",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 878.4430",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "General, Plastic Surgery",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}