Darwin Knee System
by Johnson & Johnson Professionals, Inc. 510(k)
Active
510(k) K943462
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SN | NumberK943462 | Date21 Dec 1994 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1077-2014
Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.
-
- Terminated Z-1686-2009
There may be a crack on the lateral side of the condyle in the posterior chamfer region. The device is used as a Orthopedic knee implant.
-
- Terminated Z-0403-2008
- Terminated Z-0407-2008
- Terminated Z-0402-2008
- Terminated Z-0404-2008
- Terminated Z-0405-2008
- Terminated Z-0406-2008
An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Specification
- 510(k) Number
- FDA Product Code
- JWH
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3560
- Advisory Committee
- Orthopedic
- FDA Decision
- SN
Other devices from Johnson & Johnson Professionals, Inc.
See all devices from Johnson & Johnson Professionals, Inc. →