Dca 2000 System for Hemoglobin A1C

by Heraeus Kulzer, Inc. 510(k)

Active United States Open recall

510(k) K951361

Regulatory status

as of 28 Oct 2026

Recall ongoing. 1 ongoing FDA recall names this device. See recalls

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK951361 Date5 May 1995ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

1 recall · 1 ongoing · FDA

    • Ongoing Z-1304-2024 FDA Class II initiated 7 Feb 2024

    The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.

Specification

510(k)

510(k) Number
FDA Product Code
LCP
Device Class
Class II
Regulation
21 CFR Part 864.7470
Advisory Committee
Hematology
FDA Decision
SESE Substantially Equivalent