Dca 2000 System for Hemoglobin A1C
by Heraeus Kulzer, Inc. 510(k)
Active Open recall
510(k) K951361
Regulatory status
Recall ongoing. 1 ongoing FDA recall names this device. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK951361 | Date5 May 1995 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1304-2024
The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.
Specification
- 510(k) Number
- FDA Product Code
- LCP
- Device Class
- Class II
- Regulation
- 21 CFR Part 864.7470
- Advisory Committee
- Hematology
- FDA Decision
- SESE Substantially Equivalent